| Model Number |
82009 |
| Medical Device Problem Codes |
Overheating of Device (1437); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
Intermittent vibration and overheating of the tip.No problem for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Failure mode: overheating insert root cause: tip clogged conclusion code: unexpected failure.Returned qty1 insert, 82009, 22320-00087078, 50, beyond useful life test method / gp005-wi02 inductance: gauge id# 5349 due: 11/30/2024 result: 3.12.Meets spec does not meet spec n/a.Tip condition: - meets spec does not meet spec n/a.Stack condition meets spec does not meet spec n/a.Tested on: g136 gage id# 9626-2 due date: 03/31/2023 set pressure: 35 psi.Water flow: output rate: 35 psi - meets spec does not meet spec n/a.Spray pattern: - meets spec does not meet spec n/a.Water leak: hp sealing ring proximal ring distal ring.Seam leak n/a.Vibration: - meets spec does not meet spec n/a.Grip condition: meets spec does not meet spec n/a.Other visual observation: insert is clogged, no water flow.Alleged event: confirmed - not confirmed.Additional: insert was unclogged.Insert has water flow and is now working well.Foreign matter.Insert care in the field is unknown.An inserts left standing wet may corrode or clog.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a 30k cav.Thinsert-fg,packed allegedly got hot during use.No injury occurred.
|
| |
|