• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC 30K CAV.THINSERT-FG,PACKED; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC 30K CAV.THINSERT-FG,PACKED; SCALER, ULTRASONIC Back to Search Results
Model Number 82009
Medical Device Problem Codes Overheating of Device (1437); No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event or Problem Description
Intermittent vibration and overheating of the tip.No problem for the patient.
 
Additional Manufacturer Narrative
Failure mode: overheating insert root cause: tip clogged conclusion code: unexpected failure.Returned qty1 insert, 82009, 22320-00087078, 50, beyond useful life test method / gp005-wi02 inductance: gauge id# 5349 due: 11/30/2024 result: 3.12.Meets spec does not meet spec n/a.Tip condition: - meets spec does not meet spec n/a.Stack condition meets spec does not meet spec n/a.Tested on: g136 gage id# 9626-2 due date: 03/31/2023 set pressure: 35 psi.Water flow: output rate: 35 psi - meets spec does not meet spec n/a.Spray pattern: - meets spec does not meet spec n/a.Water leak: hp sealing ring proximal ring distal ring.Seam leak n/a.Vibration: - meets spec does not meet spec n/a.Grip condition: meets spec does not meet spec n/a.Other visual observation: insert is clogged, no water flow.Alleged event: confirmed - not confirmed.Additional: insert was unclogged.Insert has water flow and is now working well.Foreign matter.Insert care in the field is unknown.An inserts left standing wet may corrode or clog.
 
Event or Problem Description
In this event it is reported that a 30k cav.Thinsert-fg,packed allegedly got hot during use.No injury occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K CAV.THINSERT-FG,PACKED
Common Device Name
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key16643634
Report Number2424472-2023-00364
Device Sequence Number17491054
Product Code ELC
UDI-Device IdentifierD003820091
UDI-PublicD003820091
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K052334
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number82009
Device Catalogue Number82009
Device Lot Number00087078
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date03/21/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2023
Supplement Date Received by Manufacturer03/21/2023
Initial Report FDA Received Date03/30/2023
Supplement Report FDA Received Date06/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-