Model Number 2K41-37
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported falsely elevated architect stat troponin-i results on four patients.Results provided: sid date initial / repeat lot# (b)(6).(b)(6).0.029 / <0.010 / <0.010 ng/ml 32677un22.(b)(6).(b)(6).0.035 / <0.010 / 0.024 ng/ml 32677un22.(b)(6).(b)(6).0.038 / <0.010 / <0.010 ng/ml 32677un22.(b)(6).(b)(6).0.031 / 0.02 / <0.010 ng/ml 32677un22.Normal range: 0.000 - 0.028 ng/ml.No impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.Patient identifier sids: (b)(6).
 
Additional Manufacturer Narrative
The complaint investigation for falsely elevated architect stat troponin-i results included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.Trending review did not identify any trends for the issue for the product.Device history record review did not identify any non-conformances or deviations with lot number 32677un22 and the complaint issue.Labeling was reviewed and sufficiently addresses the customer¿s issue.The overall performance of architect stat troponin-i was reviewed using field data from customers worldwide.The median patient population result for lot 32677un22 is comparable with all other lots in the field and confirms no systemic issue for the lots.Based on our investigation, no systemic issue or deficiency with the architect stat troponin-i reagent for lot 32677un22 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT STAT TROPONIN-I REAGENT KIT
Common Device Name
IMMUNOASSAY METHOD, TROPONIN SUBUNIT
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
Manufacturer (Section G)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key16644459
Report Number1415939-2023-00019
Device Sequence Number20206431
Product Code MMI
UDI-Device Identifier00380740052836
UDI-Public00380740052836
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K041192
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model Number2K41-37
Device Catalogue Number02K41-37
Device Lot Number32677UN22
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/10/2023
Supplement Date Received by Manufacturer04/05/2023
Initial Report FDA Received Date03/30/2023
Supplement Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/13/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2000SR INST, 03M74-02, (B)(6); ARC I2000SR INST, 03M74-02, (B)(6)
Patient SexUnknown
-
-