MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER Back to Search Results
Model Number 996081
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the gauge broke off on eagle inflation device from x-force balloon kit.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned for evaluation and further investigation was not conclusive.Though a specific cause cannot be determined, potential root causes for this event could be, "inadequate specifications" or "inadequate supplier".The device was used for treatment purposes.It is unknown if the device had met relevant specifications or resulted in the reported event.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "before use, inspect the device to verify that no damage has occurred during shipping and handling." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the gauge broke off on eagle inflation device from x-force balloon kit.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned for evaluation and further investigation was not conclusive.Though a specific cause cannot be determined, potential root causes for this event could be, inadequate specifications or inadequate supplier.The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be inappropriate package design.However, there was insufficient information to confirm this potential root cause.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: before use, inspect the device to verify that no damage has occurred during shipping and handling.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the gauge broke off on eagle inflation device from x-force balloon kit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER
Common Device Name
BALLOON DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16644552
Report Number1018233-2023-02173
Device Sequence Number6157630
Product Code LJE
UDI-Device Identifier10801741127653
UDI-Public(01)10801741127653(11)220614(17)250430(10)BMGSAT06
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K063632
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model Number996081
Device Catalogue Number996081
Device Lot NumberBMGSAT06
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2023
Supplement Date Received by Manufacturer06/26/2023
08/17/2024
Initial Report FDA Received Date03/30/2023
Supplement Report FDA Received Date07/06/2023
08/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-