MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL HD ARTHROSCOPE/SINUSCOPE, EP, 2.7MM, 30 DEG, 75MM (STORZ STYLE); RIGID ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL HD ARTHROSCOPE/SINUSCOPE, EP, 2.7MM, 30 DEG, 75MM (STORZ STYLE); RIGID ENDOSCOPE Back to Search Results
Model Number 242056
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).E3: the initial reporter is a j&j employee.H4: the device manufacture date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported through in-house distribution that during inspection, it was observed that the hd arthroscope/sinuscope, ep, 2.7mm, 30 deg, 75mm (storz style) device was cloudy.There was no procedure nor patient involvement reported.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at the manufacturing site and evaluated.During the service evaluation the following defects were identified: outer tube damaged.Distal tip damaged, distal tip fiber damaged.Optical system, optical components - distal cover glass/negative damaged.Broken (2+ pieces) : device fractured - optical.Functional : damaged connector - tube.Per service reports, this complaint can be confirmed.The defective parts needs to be replaced to resolve the issues.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD ARTHROSCOPE/SINUSCOPE, EP, 2.7MM, 30 DEG, 75MM (STORZ STYLE)
Common Device Name
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle MA CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16648545
Report Number1221934-2023-01409
Device Sequence Number20212922
Product Code HRX
UDI-Device Identifier10886705028146
UDI-Public10886705028146
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K080560
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number242056
Device Catalogue Number242056
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2023
Supplement Date Received by Manufacturer05/08/2023
Initial Report FDA Received Date03/30/2023
Supplement Report FDA Received Date05/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-