Model Number 02.124.416
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2023 during a procedure there was a inter-prosthetic fracture with a distal femur fracture.A frn was inserted and then two plates were inserted distally.A lateral plate was inserted and the same plate, but contralateral side was put on the medial side.During the frn, the proximal portion became snagged up.The drive cap broke during malleating.While inserting both plates distally, the center hole did not lock into the plate.Unsure if it was due to the traffic of the screws, and making the screws skive off axis.Depth gauge was also bent in the process of measuring for screws.There were four (4) minutes of surgical delay.This complaint involves five (5) device.This report is for one (1) 4.5 va-lcp curved cond pl/16 hole/336/rt this is report 2 of 5 for complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional narrative: additional pro-codes: hrs, hwc.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 VA-LCP CURVED COND PL/16 HOLE/336/RT
Common Device Name
CONDYLAR PLATE FIXATION IMPLANT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16649229
Report Number8030965-2023-03898
Device Sequence Number15003617
Product Code JDP
UDI-Device Identifier10886982042934
UDI-Public(01)10886982042934
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.124.416
Device Catalogue Number02.124.416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/13/2023
Initial Report FDA Received Date03/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4.5 VA-LCP CURVED COND PL/6 HOLE/159/LFT; CONNEC F/INSERTION HANDLE F/PFNA; CONNEC F/INSERTION HANDLE F/PFNA; DEPTH GAUGE FOR LCKNG SCRS TO 100
Patient SexUnknown
-
-