|
Siemens has completed a detailed technical investigation of the issue which revealed a device malfunction.The order of the orientation labels "head", "center", and "feet" in the axial thick slice segments are erroneously swapped in the following two cases: if the recon parameter "image order" is changed by the user from the default setting "head to feet" to "feet to head", or if the recon parameter "mirroring" is changed by the user from the default setting "none" to any other value, this influences the parameter "image order", which is switched from the default setting "head to feet" to "feet to head" so that the abovementioned condition applies.As a result, the head segment may mistakenly be labeled as "feet" and vice versa.If the user does not use anatomic landmarks and/or the shown images, and only focuses on the wrong orientation labels to determine the direction in which the needle must be tilted (towards the head or towards the feet), this could lead to the incorrect placement and angulation of the needle, resulting in potential unwanted damage to sensitive anatomical structures.In addition, if the "move to selected axial slice position" functionality is used to position the patient table for the next scan, and the user's selection of segment is based on the incorrect orientation label, this could lead to a scan being performed at the wrong position.To avoid a potential risk the customer should set the recon parameter "image order" to head to feet".Currently, siemens is working on a permanent solution for this malfunction and the solution will be distributed as a software update.As an immediate action, siemens has sent a customer safety advisory notice to address this to all users with the affected application licenses.The csan was reported according to 21cfr806 (fda# 2240869-12/22/2022-0021-c/ res# 91481).The affected customer received the csan on january 19, 2023 and was aware of the malfunction.
|