| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Device Handling Problem (3265)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the patients head was out of position during lgk treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed by conducting a thorough evaluation of the reported information.The customer informed that the patient became dislodged from the head frame.The hospital staff had not noticed any movement of the patient during the treatment, however all four pins from the frame were at a different location.No injury of the patient has been reported by the customer but the treatment had to be re-started with a new fixation.Movement of the patient in the gamma knife resulted in the patient's skull being outside the radiation focus.It is uncertain how much radiation the patient missed in the target since it is unknown when the patient moved.No ct or cbct images have been received from the customer to verify the frame position as well as the radiation target.There is no reason to suspect that there is something wrong with the vantage equipment and no damages could be seen on the firmfix used.The root cause is identified to be use error where the firmfix was fixated with too low torque and probably high position of the frame on the head as the patient did not have any lacerations of the skin but this has not been fully verified.The customer has been informed regarding the difference in torque between the g-frame (used by customer since 2005) and the vantage frame and has received the white paper regarding this (leksell vantage stereotactic system head frame fixation - (b)(6)).
|
| |
|
Search Alerts/Recalls
|