• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL RUSCH GREENLED HANDLE: MEDIUM; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL RUSCH GREENLED HANDLE: MEDIUM; LARYNGOSCOPE, RIGID Back to Search Results
Model Number IPN915489
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during testing, the blade cannot be removed from the handle anymore, it is stuck.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during testing, the blade cannot be removed from the handle anymore, it is stuck.No patient involvement.
 
Additional Manufacturer Narrative
(b)(4).The actual device was returned and sent to the manufacturing site for investigation.The manufacturing site reports: "the customer returned one 133300 rusch greenled handle: medium for investigation.The handle was returned with a dahlhauseen blade 3 attached to it.The visual examination did not reveal any obvious defects or anomalies, as the handle device appears typical.The blade was stuck on the handle from the blade hinge, not from the steel ball side.The blade was able to be removed with ordinary effort.All edges of the returned blade and handle were examined, and it was noted that there were scratches observed on dahlhauseen blade 3 near the hinge (slot) profile.The functional inspection was performed per laryngoscope ifu by pushing downward on the fresh blade to release it from the handle, part of the instructions for use (ifu) were reviewed for this investigation.The handle device was functional tested with fresh blade by following the product guidelines.No problem found on customer returned handle device.The device history record for lot 23012701 was reviewed and no issue that could have contributed to the reported failure was noted.The device was ma nufactured according to release specification.The devices were tested 100% for functional testing prior to ship to customer." the complaint cannot be confirmed.The device was manufactured according to release specifications and tested 100% for functional testing prior to shipment to the customer.No problem was found with the returned handle.The device was found to be fully functional with different size fiber optic blades.Other remarks: n/a.Corrected data: n/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUSCH GREENLED HANDLE: MEDIUM
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
TRUPHATEK INTERNATIONAL LTD.
14 benny gaon street
p.o. 8051
netanya 42504 43
IS   4250443
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16655924
Report Number8030121-2023-00011
Device Sequence Number17796141
Product Code CCW
UDI-Device Identifier14026704711694
UDI-Public14026704711694
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN915489
Device Catalogue Number133300
Device Lot Number23012701
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2023
Supplement Date Received by Manufacturer05/16/2023
Initial Report FDA Received Date03/31/2023
Supplement Report FDA Received Date05/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
Patient SexUnknown
-
-