• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE EXTENDED LENGTH SIDEPORTS BILIARY DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE EXTENDED LENGTH SIDEPORTS BILIARY DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number N/A
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Common device name- (additional common names and procodes): lje catheter, nephrostomy; gbo catheter, nephrostomy, general & plastic surgery.Occupation: interventional radiology lead pma/510(k) #: k173035 this report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Sometime during the week of (b)(6) 2023, an unknown patient required an ultrathane extended length sideports biliary drainage catheter.It was reported that the blue flexible stiffener was difficult to advance into the drainage catheter and then could not be removed once it was inserted into the drain.The drain and stiffener were removed, and a second drain was opened, encountering the same issue.It was not stated how the procedure was completed, but no adverse events were reported for the patient.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: on 27mar2023, cook medical inc.Received a complaint from a representative at the (b)(6) hospital.It was reported that when advancing the flexible stiffener into the ultrathane extended length sideports biliary drainage catheter (rpn: ult10.2-38-40-p-clbs-26-rh; lot: 52083313), resistance was encountered.Upon the attempted removal of the flexible stiffener, difficulty was also encountered.The drain was removed, and a second drain was opened with the same issue.No adverse effects were reported to the patient.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, quality control procedures, and specifications of the device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook reviewed the device history record (dhr).The dhr for lot 15208313 and all related subassembly lots confirmed zero non-conformances relevant to the reported difficulty.To date, a further search of our database records discovered no additional complaints received from the provided lot.Cook also reviewed the product labeling.The product ifu, [t_multi2_rev1] ¿multipurpose drainage catheter,¿ states: "precautions: when inserting a stiffening cannula into a catheter with retention suture, hold suture during cannula insertion to avoid bunching or tangling of suture.How supplied: supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ the information provided upon review of the dmr, ifu and dhr suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, no device return, and the results of the investigation, cook medical has concluded that component failure unrelated to manufacturing or design deficiencies contributed to this incident.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE EXTENDED LENGTH SIDEPORTS BILIARY DRAINAGE CATHETER
Common Device Name
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key16656417
Report Number1820334-2023-00354
Device Sequence Number15104922
Product Code GBO
UDI-Device Identifier00827002133532
UDI-Public(01)00827002133532(17)260206(10)15208313
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/06/2026
Device Model NumberN/A
Device Catalogue NumberULT10.2-38-40-P-CLBS-26-RH
Device Lot Number15208313
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/27/2023
Supplement Date Received by Manufacturer07/31/2023
Initial Report FDA Received Date03/31/2023
Supplement Report FDA Received Date08/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-