Model Number DBP-125SOLID200
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The oad was returned to csi for analysis.There were no abnormalities or damage identified with the driveshaft or crown that would have contributed to the event.There was overloading identified on the driveshaft at the lap weld and extending 1 centimeter distally.This damage prevented a test wire from passing through the device.It was unknown if the overloaded driveshaft was related to the reported event.Review of the device data log did not identify any stall events during the procedure.The root cause of the overloaded driveshaft is unable to be conclusively determined.When tested, the oad functioned as intended.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Event or Problem Description
During the orbital atherectomy procedure, the stealth 360 orbital atherectomy device (oad) prolapsed near the patient's heart when it was introduced.The physician re-obtained access via the patient's other arm rather than via pedal access.A new oad with greater length was used and the same event occurred.Under fluoroscopy, it was observed the oad was not advancing.The physician observed the oad prolapsed in the heart once again.Balloon angioplasty was used to complete the procedure.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
aaron stevens
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key16656799
Report Number3004742232-2023-00090
Device Sequence Number7263567
Product Code MCW
UDI-Device Identifier10850000491219
UDI-Public(01)10850000491219(17)230430(10)379387-1
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model NumberDBP-125SOLID200
Device Catalogue Number7-10057-05
Device Lot Number379387-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2023
Initial Report FDA Received Date03/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
-
-