| Catalog Number |
07027885190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample was requested for a sample interference investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys t4 assay results for 1 patient sample on a cobas 6000 e601 module and a cobas e 801 analytical unit.This medwatch will cover t4.Refer to medwatch with a1 patient identifier (b)(6).For information on the ft3 iii results.The t4 result on the e601 analyzer was >320 nmol/l.The sample was tested on another laboratory's e801 analyzer and the result was 186 nmol/l.No questionable results were reported outside of the laboratory.The e601 analyzer serial number (b)(4).The other laboratory's e801 analyzer serial number was requested but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Two patient samples were provided for investigation.An interfering substance to the ruthenium component of the reagent was detected.Per product labeling, "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design." the investigation did not identify a product problem.
|
| |
|