| Model Number |
IPN038077 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during insertion, two catheters consecutively would not fit through the sheath.As a result, a third catheter was used and was successful.All catheters were inserted at the same insertion site.No patient harm or injury.The patient status is reported as "fine".See associated mdr 3010532612-2023-00195.
|
| |
|
Event or Problem Description
|
|
It was reported that during insertion, two catheters consecutively would not fit through the sheath.As a result, a third catheter was used and was successful.All catheters were inserted at the same insertion site.No patient harm or injury.The patient status is reported as "fine".See associated mdr 3010532612-2023-00195.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The serial number (b)(6) reported by the customer is an invalid serial number.The serial number of the returned sample is my21190.The lot number (18f22l0046) reported matches the lot number for the returned sample.Returned for investigation was a 40cc 8.0fr rediguard intra-aortic balloon catheter (iabc) without the original packaging.Upon return, the peel away sheath was noted on the returned iabc.The one-way valve was connected and tethered to the short driveline tubing.The iabc bladder was fully unwrapped.A kink to the iabc central lumen was noted near the iabc bifurcate at approximately 70.2cm from the iabc distal tip.A bend to the iabc central lumen was noted at approximately 4.4cm from the iabc distal tip.Dried blood was noted on the exterior surfaces of the returned iabc; no blood was noted within the helium pathway.The sheath in question was not returned with the sample.A handwritten note was returned with the sample stating "balloon unwrapped upon opening package , catheter would not go through sheath".The bladder thickness was measured at six points with measurements ranging from 0.0073in-0.0077in and was within specification.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute and then 30 seconds five separate times.The iabc central lumen was aspirated and flushed using a 60cc lab-inventory syringe.Some blood was noted.The iabc was leak tested and no leaks were detected.Full inflation was achieved.The device passed the leak test.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.Resistance was noted at approximately 2.4cm, 4.3cm and 70.1cm from the iabc distal tip.The guidewire was able to advance through the central lumen.Some blood was noted.The guidewire was front loaded through the iabc luer.Resistance was noted at approximately 14.5cm, 80.2cm and 82.2cm from the iabc luer.The guidewire was able to advance through the central lumen.Some blood was noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint of iab unwrapped prematurely is confirmed.The iabc bladder was fully unwrapped upon receipt of the sample.The unwrapped bladder can result in difficulty advance the iabc into the sheath/patient.The sheath in question was not returned with the sample.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the unwrapped bladder.The root cause of the complaint is undetermined.No further action required at this time.This will be monitored for any developing trends.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Search Alerts/Recalls
|