VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number V+O+C+S+N, JAPANESE
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to ventec that the ventilator measured lower peep (positive end expiratory pressure) than set.There were no reports of patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
Ventec will perform an evaluation of the device.A follow-up report will be submitted when the investigation is complete as defined by 21 cfr 803.56.
 
Additional Manufacturer Narrative
H6: the device was evaluated by ventec where the reported issue of lower peep (positive end expiratory pressure) than set was confirmed.Ventec replaced the internal flow transducer (ift) to resolve the reported issue.Proper device operation was observed through functional and performance testing.The investigation determined that the root cause of the reported issue was the ift.
 
Additional Manufacturer Narrative
Corrected information for supplemental medwatch report 002 ¿.Section d4, model #, of the initial medwatch report stated: prt-01100-100.Section d4, model #, of the initial medwatch report should have stated: v+o+c+s+n, japanese.Section d4, udi #, of the initial medwatch report stated: (b)(4).Section d4, udi #, of the initial medwatch report should have stated: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer Contact
mike kubany
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key16660733
Report Number3013095415-2023-00245
Device Sequence Number1700338
Product Code CBK
UDI-Device Identifier00850018761130
UDI-Public(01)00850018761130(11)210309
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K162877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV+O+C+S+N, JAPANESE
Device Catalogue NumberPRT-01100-100
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2023
Supplement Date Received by Manufacturer06/19/2023
11/07/2025
Initial Report FDA Received Date04/01/2023
Supplement Report FDA Received Date07/10/2023
12/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-