• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD CARIBE LTD. BD MICROTAINER® SST¿ - AMBER; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD CARIBE LTD. BD MICROTAINER® SST¿ - AMBER; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 365978
Medical Device Problem Code Device Ingredient or Reagent Problem (2910)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that prior to use with bd microtainer® sst¿ - amber had abnormal additive, it appeared "melted/sticky".The following information was provided by the initial reporter: customer states coating in the middle of the tube is melted/sticky.
 
Additional Manufacturer Narrative
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, 50 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to deformed tubes (melted/ sticky) as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode deformed tubes (melted/ sticky).Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that prior to use with bd microtainer® sst¿ - amber had abnormal additive, it appeared "melted/sticky".The following information was provided by the initial reporter: customer states coating in the middle of the tube is melted/sticky.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® SST¿ - AMBER
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
Manufacturer (Section G)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16661104
Report Number2618282-2023-00021
Device Sequence Number6616628
Product Code JKA
UDI-Device Identifier50382903659789
UDI-Public(01)50382903659789
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K991702
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2023
Device Model Number365978
Device Catalogue Number365978
Device Lot Number2285923
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2023
Supplement Date Received by Manufacturer04/14/2023
Initial Report FDA Received Date04/02/2023
Supplement Report FDA Received Date04/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-