COVIDIEN ABRE VENOUS SELF-EXPANDING STENT SYSTEM; STENT, ILIAC VEIN Back to Search Results
Model Number AB9U12060090
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 03/07/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Two abre stents self-expanding were implanted for the treatment of a fibrous lesion in the left common iliac vein and left external iliac vein.There was mild tortuosity and no calcification.The devices were prepped as per the ifu with no issues identified.The lesion was pre-dilated.The procedure was completed with no issues mentioned.During a follow up visit the stents were noted to be occluded.Reintervention on this patient with the doctor on (b)(6) 2023.During the case, the stents were reopened the by removing the present clot with aspirational thrombectomy and using a high-pressure balloon.The stents were re-evaluated that were originally placed, and the physician had the following conclusions: unsure about why the stents went down.Stents appeared to be properly sized according to ivus ¿ no space between vessel wall and stent, no ¿step¿ into native vessel which would mean an oversizing.After clearing stents of clot and reopening, no collateral vessels appeared on venogram.This led the doctor to believe that there are no vessels that are stealing flow from the iliac veins after the case no clinical reason could be confirmed why the stents shut down at that time.Physician chose to not place any additional stents.No further patient injury reported.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABRE VENOUS SELF-EXPANDING STENT SYSTEM
Common Device Name
STENT, ILIAC VEIN
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16662501
Report Number2183870-2023-00110
Device Sequence Number12413868
Product Code QAN
UDI-Device Identifier00643169796225
UDI-Public00643169796225
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P200026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/29/2025
Device Model NumberAB9U12060090
Device Catalogue NumberAB9U12060090
Device Lot NumberB454696
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2023
Supplement Date Received by Manufacturer04/12/2023
Initial Report FDA Received Date04/03/2023
Supplement Report FDA Received Date04/17/2023
Date Device Manufactured09/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age44 YR
Patient SexFemale
-
-