• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH BLUNT TIP SCREW, 4X48MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH BLUNT TIP SCREW, 4X48MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 47-2486-048-40
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/21/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a checkup, approximately 4 weeks after implantation, the surgeon noticed that one of the proximal screws had backed out from its proper position.No revision surgery has been scheduled.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).Medical devices: proximal humerus, right, 9x160mm; item# 47249616009; lot# 3082274.Blunt tip screw, 4x40mm; item# 47248604040; lot# 3091511.Blunt tip screw, 4x42mm; item# 47248604240; lot# 3126120.Report source- foreign: japan.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2023-00145, 0009613350-2023-00147, 0009613350-2023-00148.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Remains implanted.
 
Additional Manufacturer Narrative
(b)(4).No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Medical records were not provided.Based on the investigation, it could can be assumed that possible contributing factors contributing to the migration of the screws might could be multifactorial related to either patient condition, behavior or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the locking mechanism of the nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUNT TIP SCREW, 4X48MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key16663034
Report Number0009613350-2023-00149
Device Sequence Number6607133
Product Code HSB
UDI-Device Identifier00889024505483
UDI-Public(01)00889024505483(17)260917(10)3087851
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181827
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/17/2026
Device Model Number47-2486-048-40
Device Catalogue Number47248604840
Device Lot Number3087851
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2023
Supplement Date Received by Manufacturer04/25/2023
Initial Report FDA Received Date04/03/2023
Supplement Report FDA Received Date05/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexPrefer Not To Disclose
Patient Weight74 KG
-
-