| Model Number |
2545-00-672 |
| Medical Device Problem Codes |
Break (1069); Crack (1135); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was damaged.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.Review of the photographic evidence revealed that the trial exhibited scratches and nicks covering a large portion of the surface and around the edges.However, no sings of breakage nor cracks were found, therefore the reported allegations can be partially confirmed. superior edge of the trial was chipped.Additionally, scratches and nicks can be observed on the surface and around the edges.However, no sings of cracks were found.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) investigation summary - the device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned sample revealed that the device was severely chipped, exhibited scratches and nicks covering a large portion of the surface.Additionally, a crack was found around one of the metal inserts, therefore the complaint can be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = > the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.Review of the photographic evidence revealed, that the trial exhibited scratches.And nicks covering a large portion of the surface and around the edges.However, no signs of breakage nor cracks were found.Therefore, the reported allegations can be partially confirmed. superior edge of the trial was chipped.Additionally, scratches and nicks can be observed, on the surface and around the edges.However, no signs of cracks were found.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|