• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG DEVICE, SUBCUANEOUS FACIA CLOSURE; BERCI FASCIAL CLOSURE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG DEVICE, SUBCUANEOUS FACIA CLOSURE; BERCI FASCIAL CLOSURE DEVICE Back to Search Results
Model Number 26173AM
Medical Device Problem Codes Material Separation (1562); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was event with a 26173am a fascial closure instrument 2.8mm, 17cm.According to the information received via the medwatch report 0501290000-2023-8001 and the facility, during a laparoscopic appendectomy procedure, at closing the closure device broke outside of the patient on the sterile field.Per the facility no patient harm and procedure was completed.
 
Additional Manufacturer Narrative
Manufacturing site evaluation: the instrument evidence of use: the bottom jaw was broken off and the outer tension rod was found on the front of closure handle instead of the rear.In addition, the shaft was found to be bent.The event is filed under internal karl storz complaint id (b)(4).The ifu pi-000052 ifu v6.0 states "inspect the instrument for cleanliness and damage.Do not use instruments that show visible signs of damage or that are difficult to use.Any malfunction of an instrument during a procedure could result in injury to the patient and further damage to the instrument.".
 
Additional Manufacturer Narrative
Manufacturing site evaluation: the fracture of the jaw is due to overstressing.The high age (nov.2013) of the product and the associated reconditioning cycles may have favored this damage pattern.The event is filed under internal karl storz complaint id (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEVICE, SUBCUANEOUS FACIA CLOSURE
Common Device Name
BERCI FASCIAL CLOSURE DEVICE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen, gm
Manufacturer (Section G)
KARL STORZ SE & CO.KG
dr.-karl-storz -strasse 34
78532
tuttlingen, gm,
GM  
Manufacturer Contact
anja fair
2151 e. grand avenue
el segundo, CA 90245
4242188247
MDR Report Key16664408
Report Number9610617-2023-00069
Device Sequence Number17772541
Product Code GCJ
UDI-Device Identifier04048551097755
UDI-Public4048551097755
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K942612
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26173AM
Device Catalogue Number26173AM
Device Lot NumberYW01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2023
Supplement Date Received by Manufacturer07/07/2023
Initial Report FDA Received Date04/03/2023
Supplement Report FDA Received Date09/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age33 YR
Patient SexFemale
Patient Weight73 KG
-
-