| Model Number |
90620US |
| Medical Device Problem Code |
Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported that the cord leading to the battery pack of the duo headlight 2 bay system - us (90620us) has a short in it.No patient/user injury, death or surgical delay occurred as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported duo headlight 2 bay system - us (90620us) was not returned for investigation; therefore, an evaluation of the device could not be performed and a definite root cause of the reported issue could not be identified.Based on the reported complaint, the probable root cause for this 90620us headlight having a short in the cord from the headband to the battery pack is damage to the cord caused by rough handling/environmental damage.The issue of a "short" refers to poor connection of the power pins that can potentially result in a failing connection to the battery port.No risk of harm would be incurred by this type of short, and the low voltage would not cause a hazard.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
| |
|