| Catalog Number |
03.632.001 |
| Medical Device Problem Code |
Compatibility Problem (2960)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device report from synthes reports an event in japan as follows: it was reported that this was an interbody fusion (l3, left) for stenosis on (b)(6)2023.The surgeon inserted the screw and tried to detach it from the holding sleeve, but it did not come off.The surgeon tried various methods, but the screw did not come off at all.The sales rep asked the surgeon to remove the screw once.Then, the surgeon inserted another screw using another holding sleeve.The inserted screw was detached from the holding sleeve with no problem.The defective holding sleeve could not be detached from the screw during surgery.When the sales rep tried to detach the holding screw from the screw after surgery, it did not come off at all with the usual detaching method, but the sale rep was finally able to detach them by repeated pulling, etc.After several attempts to detach and attach the holding sleeve and the screw, the screw would not come off the holding sleeve about 1 out of every 3 times.No further information is available.This report is for one (1) retain-sleeve std f/matrix 5.5 this is report 1 of 1 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Part: 03.632.001.Synthes lot: h364388.Supplier lot:h364388.Release to warehouse date: april 12, 2018.Manufactured by: avalign technologies-nemcomed.No ncrs were generated during production.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the retain sleeve and polyaxial screw were returned in a disassembled condition.It was observed that the threaded end of the retain sleeve was heavily stripped.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.A functional test revealed the retain sleeve was able to assemble to the polyaxial head of the screw.After that, the devices were disassembled without any sign of resistance.The complaint condition was not replicated.After a visual inspection per proceudre, it was determined that the reusable instrument device was stripped from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed stripped of retain-sleeve std f/matrix 5.5, p/n: 03.632.001, lot: h364388 would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|