| Model Number |
PED2-475-16 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that two pipelines failed to open in the distal section.The patient was undergoing surgery for treatment of a fusiform, ruptured aneurysm in the supraclinoid internal carotid artery (ica) with a max diameter of 7 mm and a 7 mm neck diameter.The landing zone was 5 mm distally and 4.5 mm proximally.It was noted the patient's vessel tortuosity was severe.It is unknown if dual antiplatelet treatment was administered.The angiographic result post procedure is unknown.It was reported that it was attempted to open both devices in m1 segment and recovered wings.Both devices did not open even when pushed out of microcatheter more than 50%.It was reported there was failure to open in the distal section for both pipelines.The pipeline was not positioned in a bend.More than 50% of the pipeline had been deployed when it failed to open.The pipeline was resheathed less than or equal to two times.There were no additional steps or other devices required to open the pipeline.The pipeline was removed from the patient by placing the microcatheter and removing from the body.The pipeline was not resheathed and removed with the microcatheter.The second pipeline with lot # b414565 was removed from the patient by resheathing and removing with the microcatheter.The pipeline was used for an indication that is off-label, specifically use for supraclinoid internal carotid artery.The pipeline and any accessory devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or further complications were reported as a result of this event.Ancillary devices include a sophia 6f guide catheter and phenom 27 microcatheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the cause of the pipeline failure to open was not determined.There was no damage observed to the pipeline after removal from the patient.
|
| |
|