Model Number HUM LNR IMPACT TIP 36MM
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The center of the poly impactor broke while impacting the poly.There was a 30 seconds surgical delay which was enough time to switch to the 38mm impactor.No patient impact reported due to the delay.Patient was last known to be in stable condition following the event.The product was returned for analysis.
 
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
H6: based on historical complaint investigations, the image provided, and the reported event, the fractured humeral liner impactor tip reported was likely the result of a stress riser introduced during impaction, which led to crack initiation, propagation, and ultimate fracture.Additionally, the device may have experienced material degradation if sterilized beyond the recommended parameters listed in the reprocessing instruction which could have led to the observed failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key16667660
Report Number1038671-2023-00568
Device Sequence Number6501251
Product Code LXH
UDI-Device Identifier10885862558831
UDI-Public10885862558831
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHUM LNR IMPACT TIP 36MM
Device Catalogue Number321-09-36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2023
Supplement Date Received by Manufacturer06/26/2023
05/11/2026
Initial Report FDA Received Date04/03/2023
Supplement Report FDA Received Date06/26/2023
05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-