| Model Number |
HUM LNR IMPACT TIP 36MM |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The center of the poly impactor broke while impacting the poly.There was a 30 seconds surgical delay which was enough time to switch to the 38mm impactor.No patient impact reported due to the delay.Patient was last known to be in stable condition following the event.The product was returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: based on historical complaint investigations, the image provided, and the reported event, the fractured humeral liner impactor tip reported was likely the result of a stress riser introduced during impaction, which led to crack initiation, propagation, and ultimate fracture.Additionally, the device may have experienced material degradation if sterilized beyond the recommended parameters listed in the reprocessing instruction which could have led to the observed failure.
|
| |
|