|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The initial reporter is a j&j sales representative.Investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection, it was found that the anchor was detached from the inserter tip.The distal end of the inserter was deformed which created a problem to hold the anchor in place.A manufacturing record evaluation was performed for the finished device lot number: 9l37817, and no non-conformances were identified.Based on the condition of the device received, this complaint can be confirmed.As per ifu: do not twist or apply bending force to the inserter.Doing so may damage the anchor, suture, or inserter tip.Excessive tension may overload the anchor or suture.The possible root cause for the issue reported can be attributed to excessive force and twist applied during procedure.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
|
It was reported by the sales rep in south korea that during an unknown procedure on (b)(6) 2023, it was observed that the minilok qa+ w/ #0 ocord and os-2 needle w/bit device had an unspecified malfunction.During in-house engineering evaluation, it was determined that the anchor on the device was detached from the inserter tip.It was further determined that the distal end of the inserter on the device was slightly deformed.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|