• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD SHORTSHOT SAEED HEMORRHOIDAL MULTI-BAND LIGATOR WITH TRIVIEW ANOSCOPE; FER LIGATOR, HEMORRHOIDAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD SHORTSHOT SAEED HEMORRHOIDAL MULTI-BAND LIGATOR WITH TRIVIEW ANOSCOPE; FER LIGATOR, HEMORRHOIDAL Back to Search Results
Model Number G34063
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: the device involved in this complaint was not available to be returned, therefore a document based investigation will be complete.This file is related to (b)(4).Documents review including ifu review: prior to distribution hmbl-4-tri devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for hmbl-4-tri of lot number c1962064 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1962064.As per the instructions for use, ifu0030, users are advised of the following: "if the package is opened or damaged when received, do not use.Visually inspect with particular attention to joints, cracks or breaks.If an abnormality is detected that would prohibit proper working condition, do not use.¿ the ifu states that ¿ maintain suction of haemorrhoid by keeping suction port covered.Deploy band by slowly rotating suction spool downwards until tension is released.¿ ¿uncover suction port of ligator to relieve suction on haemorrhoid.This will allow ligated haemorrhoid toe be released from tip of device.¿ root cause review: a definitive root cause for the customer complaint could not be determined from the available information.A possible root cause could be attributed to the users technique during the procedure.It is possible that if suction was not released or too much tissue was sucked into the tip of the device, this may have caused the bands to slip off the haemorrhoid as the tissue may not have bulged around the band holding the bands in place , or if the user did not begin at the most proximal location from the anal sphincter and proceed distally resulting in passing the shortshot over a previously placed band this may dislodge the band.Several attempts have been made to request additional information but it has not been forthcoming, if this is received at a later date the investigation will be updated accordingly.Summary: complaint is confirmed based on customer testimony.The patient did not experience any adverse events due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
The rep reported a complaint from reservoir private regarding 4 x shortshot (hmbl-4-tri) failing to engage.Lot numbers: 1 x c19665 (rep asked to confirm lot) - (b)(4).1 x c1962064 - (b)(4).1 x c19634 (rep asked to confirm lot) - (b)(4).1x c1966535 - (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHORTSHOT SAEED HEMORRHOIDAL MULTI-BAND LIGATOR WITH TRIVIEW ANOSCOPE
Common Device Name
FER LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
sinead o'leary
o halloran road
national technology park
limerick 
MDR Report Key16673670
Report Number3001845648-2023-00213
Device Sequence Number5434776
Product Code FER
UDI-Device Identifier00827002340633
UDI-Public(01)00827002340633(17)230719(10)C1962064
Combination Product (Y/N)N
PMA/510(K) Number
K060623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/19/2023
Device Model NumberG34063
Device Catalogue NumberHMBL-4-TRI
Device Lot NumberC1962064
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date09/26/2022
Event Location Hospital
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/27/2022
Initial Report FDA Received Date04/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-