|
ETHICON INC. PRLNE BLU 24IN 5-0 D/A C-1 VB; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
|
Back to Search Results |
|
| Model Number |
8325H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).H6: component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Related reports: 2210968-2023-02345.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an aorto-bifemoral bypass surgery on an unknown date and suture was used.During the procedure, when the surgeon tried to make a knot with the suture, the suture got torn.The scrub nurse gave dr.(b)(6) a new suture from the same package, and again, while making a knot, even after wetting his gloves, the suture was torn.The scrub nurse gave dr.(b)(6) a new suture from a different package, and he was able to make a knot.The surgery continued and was completed successfully.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 4/25/2023.A device has been received, however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-02345.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) date sent to the fda: 5/17/2023 h6 component code: g07002 no device problem found additional information: h3, h6 h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional test evaluation were conducted on the returned device.Visual analysis of the returned sample determined that one opened box with twenty-six packets that pertain to product code 8325h was received for evaluation.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to damaged or breakage sutures and no defects were observed during evaluation.A functional test was performed, and the tensile strength results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-02344 & 2210968-2023-02345.
|
| |
|
Search Alerts/Recalls
|
|
|