MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HD CARTRIDGE LINES; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HD CARTRIDGE LINES; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 103401
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter name and address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported particulate matter was observed in the arterial chamber of a gambro cartridge line.This event was noted prior to patient use.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Additional information: d9, h3, h4, h6 and h10.H10: the device was received for evaluation containing an unspecified volume of fluid.During visual inspection, a brown/black particle was observed inside and embedded in the venous filter.The particle underwent microscopic inspection, the material was related resin/plastic degraded.The reported condition was verified.The cause of the condition was due to the molding process during manufacturing; the revolutions in which the material moves and the temperature variations, the remains of material are exposed and cause degradation in the plastic.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD CARTRIDGE LINES
Common Device Name
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TIJUANA BAJA
blvd pacifico 10014
parque industrial pacifico
tijuana baja california CP 22 643
MX   CP 22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16674961
Report Number8030638-2023-00008
Device Sequence Number17479601
Product Code FJK
UDI-Device Identifier07332414007867
UDI-Public(01)07332414007867
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K070414
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/15/2025
Device Catalogue Number103401
Device Lot NumberMA0122109575
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2023
Supplement Date Received by Manufacturer05/04/2023
Initial Report FDA Received Date04/04/2023
Supplement Report FDA Received Date05/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-