| Catalog Number |
103401 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter name and address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported particulate matter was observed in the arterial chamber of a gambro cartridge line.This event was noted prior to patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3, h4, h6 and h10.H10: the device was received for evaluation containing an unspecified volume of fluid.During visual inspection, a brown/black particle was observed inside and embedded in the venous filter.The particle underwent microscopic inspection, the material was related resin/plastic degraded.The reported condition was verified.The cause of the condition was due to the molding process during manufacturing; the revolutions in which the material moves and the temperature variations, the remains of material are exposed and cause degradation in the plastic.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|