| Catalog Number |
07.02067.001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported, the tip of a vitality straight pedicle probe broke off in the patient's vertebral body intra-operatively, while the screw hole was being prepared.The fractured piece was recovered using another instrument.And the procedure was completed without patient impacts.
|
| |
|
Event or Problem Description
|
|
It was reported the tip of a vitality straight pedicle probe broke off in the patient's vertebral body intra-operatively while the screw hole was being prepared.The fractured piece was recovered using another instrument and the procedure was completed without patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d1; d4: catalog, lot, and udi numbers; d9; and h3.Additional information in h4.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in g1 and h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed that the tip of the device was fractured off.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.This event could also be attributed to off-axis forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported the tip of a vitality straight pedicle probe broke off in the patient's vertebral body intra-operatively while the screw hole was being prepared.The fractured piece was recovered using another instrument and the procedure was completed without patient impacts.
|
| |
|