• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, ADULT; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, ADULT; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 670175
Medical Device Problem Codes Leak/Splash (1354); Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported lot number is assumed but not confirmed.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that a trach had a cuff leak that was discovered while in use for twelve (12) days with a patient.Leakage was confirmed when the product was withdrawn from the patient and further tested.No injury was reported.
 
Additional Manufacturer Narrative
One device and two photos were received for investigation.The device was visually inspected and was detected that the cuff of the trachel tube had a tear.A leak test was performed and the device failed the test.The complaint is confirmed.According to the customer's report, the product was found damaged until its use, during the pre-check no leakage was reported on the cuff according with the instructions for use, and it was until it was withdraw from patient after 12 days when it leaked, the root cause could be due to improper use of the product.A review of the device of the device history records could not be completed as the lot number provided by the customer was incorrect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, ADULT
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16677971
Report Number3012307300-2023-03520
Device Sequence Number5347347
Product Code JOH
UDI-Device Identifier15021312006063
UDI-Public15021312006063
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K944178
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number670175
Device Catalogue Number670175
Device Lot Number4257599
Was Device Available for Evaluation? No
Date Returned to Manufacturer03/20/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2023
Supplement Date Received by Manufacturer06/19/2023
Initial Report FDA Received Date04/04/2023
Supplement Report FDA Received Date06/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/18/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-