TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS Back to Search Results
Model Number 550AVHCT
Medical Device Problem Codes Thermal Decomposition of Device (1071); Electrical Power Problem (2925)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit blew a fuse so it could not run on alternating current (ac) power.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Additional Manufacturer Narrative
During laboratory evaluation, the product surveillance technician (pst) observed the monitor to not power up.The fuses were removed and both were found to be in the open state meaning there was no continuity through either fuse.He also observed the power supply to have experienced an overheating event and evidence of an electrical discharge that caused the power supply to be damaged and fail.It was determined that the monitor did not meet specification.During testing of the device at the service center, the monitor failed to power up due to the defective power supply.The power supply and its associated parts and the ac power cable were replaced, and the monitor then successfully powered on.The monitor will be brought to manufacturer's specifications before being returned to the customer.
 
Additional Manufacturer Narrative
Updated block: h6.The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 550
Common Device Name
CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key16678480
Report Number1828100-2023-00102
Device Sequence Number6379638
Product Code DRY
UDI-Device Identifier00886799001783
UDI-Public(01)00886799001783(11)190321
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K182110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number550AVHCT
Device Catalogue Number550AVHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2023
Supplement Date Received by Manufacturer04/11/2023
Initial Report FDA Received Date04/04/2023
Supplement Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-