MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET Back to Search Results
Model Number 66800039
Medical Device Problem Codes Connection Problem (2900); Noise, Audible (3273)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during an arthroscopy, the versajet ii console does not do its function, and makes a lot of noise when used with a 8-position connection, and according to the customer, it does not work with the 14-position connection.The procedure was cancelled.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference number:(b)(4) section b5 was corrected.
 
Event or Problem Description
It was reported that, during a wound debridement treatment, the versajet ii console does not work: it makes a lot of noise when used with a 8 mm handpiece, and it does not work with the 14 mm.The procedure was cancelled.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
Section h10: the device was returned for evaluation.A visual and functional assessment was performed.No material nor functional failures was observed, the device passed all testing.A documentation review was performed.The manufacturing records and inspection processes for the reported batch contain no contributory factors, records confirm that this batch was manufactured, meeting the required specifications.All inspection and testing requirement was met prior to the release.Historical review details previous cases of this nature, which are considered isolated in scope and there are no open nor closed escalation actions related to this event type.The risk management files adequately captures both of the reported concerns, with no updates required.The probable cause remains unknown.Factors may include a defect with the actual handpiece used, but that component has not bee returned for evaluation to support this case.No manufacturing, packaging, labelling, design, concerns nor adverse trend have been observed, therefore no corrective actions are deemed necessary.Internal complaint reference number: (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference number: (b)(4) section h6 (health effect - clinical code and health effect - impact code) was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET II CONSOLE
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16678573
Report Number8043484-2023-00024
Device Sequence Number5435439
Product Code FQH
UDI-Device Identifier00040565124759
UDI-Public00040565124759
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K143115
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66800039
Device Catalogue Number66800039
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2023
Supplement Date Received by Manufacturer06/23/2023
06/23/2023
09/25/2024
Initial Report FDA Received Date04/04/2023
Supplement Report FDA Received Date06/26/2023
06/27/2023
10/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-