| Model Number |
6290 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the communicator was dark and was extremely warm to touch.A boston scientific company representative discussed with the patient and opted to replace the entire system.The communicator was deactivated.A return label was sent.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory additional information from product analysis indicates that the allegation was confirmed.During visual analysis, the communicator adapter had a missing piece, and the power adapter was not snug when plugged in.The communicator was taking apart, and areas were burnt and could smell a smoky, fire smell.Resistance testing on a component was not performed due to the component was being completely burnt.The communicator behavior was consistent with a high-power electrical overstress event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the communicator was dark and was extremely warm to touch.A boston scientific company representative discussed with the patient and opted to replace the entire system.The communicator was deactivated.A return label was sent.No adverse patient effects were reported.
|
| |
|