• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFECELL 6CM X 10CM STRATTICE FIRM STOMA REINFORCED SHEET; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFECELL 6CM X 10CM STRATTICE FIRM STOMA REINFORCED SHEET; MESH, SURGICAL Back to Search Results
Catalog Number 0610008
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hernia (2240)
Date of Event 10/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported through a legal event that a 61 year old male patient had hernia repair surgery on or about (b)(6) 2015.During the hernia repair surgery, the surgeon implanted a strattice mesh.The records indicate this is a 10 cm x 6 cm strattice biologic mesh.After surgery, the patient returned to the hospital on or about (b)(6) 2021 for a revision surgery.No other information was reported.
 
Additional Manufacturer Narrative
This legal event is being reported as serious injury due to the reported recurrence with surgical intervention.The lot associated with this event was not reported and remains unknown; therefore an internal investigation into the device history records could not be performed.No strattice devices were returned for evaluation.Based on the limited information, including no identification of the lot number, and without relevant patient factors, a relationship between the event and the strattice could not be determined.Due to the legal process, if additional information is made available during legal proceedings, a supplemental report will be submitted.
 
Event or Problem Description
This is follow up #1 to report on 04/apr/2023, pmqa received notification from legal that the lot associated with this event was found through discovery and is sp100080-092.No other information was reported.As reported in the initial: it was reported through a legal event that a 61 year old male patient had hernia repair surgery on or about (b)(6) 2015.During the hernia repair surgery, the surgeon implanted a strattice mesh.The records indicate this is a 10 cm x 6 cm strattice biologic mesh.After surgery, the paitent returned to the hospital on or about (b)(6) 2021, for a revision surgery.No other information was reported.
 
Additional Manufacturer Narrative
Internal investigation into strattice lot sp100080 included a review of the reported information, review of the device history records, and a review of the complaint history records.The investigation resulted in no remarkable findings, including no other complaints reported against the lot and no deviations or related non-conformances revealed during processing.The lot was terminally sterilized within the process parameters and met all qc release criteria.As of 17 april 2023, all of the 223 devices released to finished goods for lot sp100080 have been distributed with 125 reported as implanted.Based on our internal investigation with no remarkable findings, and without relevant patient factors, a relationship between the strattice and this event could not be determined.Due to the legal process, if additional information is made available during legal proceedings, a supplemental report will be submitted.As reported in the initial: this legal event is being reported as serious injury due to the reported recurrence with surgical intervention.The lot associated with this event was not reported and remains unknown; therefore an internal investigation into the device history records could not be performed.No strattice devices were returned for evaluation.Based on the limited information, including no identification of the lot number, and without relevant patient factors, a relationship between the event and the strattice could not be determined.Due to the legal process, if additional information is made available during legal proceedings, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6CM X 10CM STRATTICE FIRM STOMA REINFORCED SHEET
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
LIFECELL
1 millennium way
branchburg NJ 08876
Manufacturer (Section G)
LIFECELL
1 millennium way
branchburg NJ 08876
Manufacturer Contact
christopher belle
1 millennium way
branchburg, NJ 08876
9089471470
MDR Report Key16681567
Report Number1000306051-2023-00079
Device Sequence Number8079841
Product Code FTM
UDI-Device Identifier00818410010041
UDI-Public00818410010041
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K070560
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2015
Device Catalogue Number0610008
Device Lot NumberSP100080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/21/2023
Supplement Date Received by Manufacturer04/04/2023
Initial Report FDA Received Date04/05/2023
Supplement Report FDA Received Date05/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/14/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NO INFORMATION REPORTED
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
-
-