| Model Number |
60-05-60 |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Livanova deutschland has received a report that the electrical remote-controlled tubing clamp (erc) is not working correctly during a procedure.There is no report of any patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the s5 erc tubing clamp.The incident occurred in united states.A livanova field service engineer was dispatched to the customer site and the erc was replaced.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|
Additional Manufacturer Narrative
|
|
Through follow-up communication livanova learned that the involved electrical remote-controlled tubing clamp (erc) was not sensing tubing and making squealing sounds when clamping and unclamping.As per device instructions for use, the issue that the tubing is not recognized it means that the clamping lever is in the ¿close¿ position.Simultaneously, an alert tone sounds at the scp system.Even in the case the clamp remains in closed position, it can always be manually opened.Automatic clamps devices are equipped with a handle aimed at the manual opening of the clamp releasing the engaged tube.Therefore, in case of undesired closure of the automatic clamps occurs, opening it is always possible to act on the handle to prompt perform a manual opening which is unlikely to cause or contribute to death or serious injury.Therefore, the event has been re-assessed as not reportable.Based on the review of similar complaints and according to device instructions for use, the "no tubing recognized" error message can be displayed when: - no tube is inserted into the clamp; - the wrong tube is inserted into the clamp.Regarding the noise issue and considering that the claimed clamp was manufactured in 2022, the wearing of the unit ball bearing can be ruled out and most likely the noise has been caused by residues of dried fluids on the shafts, which which obstruct the movements generated noise.
|
| |
|
Search Alerts/Recalls
|