|
(b)(4).Date sent: 4/5/2023.D4: batch # unk.Additional information was requested and the following was obtained: "did any pieces fall into the patient? if yes, were they retrieved? yes, it fell off but retrieved." "there was no change in procedure (e.G.Additional porthole, additional wound) when retrieving the broken piece from the chest.The patient is stable.The surgeon's comment: the intercostal was narrow so there is a possibility that tip was deformed and damaged due to the pressure.And also, i have a similar experience in the past.The broken piece will be returning with the device." investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the tt012 device was received with the tip of the sleeve broken.In addition, the packaging opened was returned along with the instrument.No conclusion could be reached regarding what may have caused the reported incident.One possible cause for the damage found may be due to excessive force being applied to the device.Please refer to the instructions for use for additional information regarding proper device usage.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
|