| Catalog Number |
08445699190 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation of questionable results for 1 patient sample tested with a1cx3 assay on a cobas pro c 503 analytical unit serial number (b)(4).The following results were reported: initial result: 10.8%, repeated result: 9.5%.The results were reported outside of the laboratory.The repeated result of 9.5% deemed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service engineer found that a1c qc was out of range.The customer replaced the calibrator with a new lot.Qc was performed and it was within the normal range.
|
| |
|
Additional Manufacturer Narrative
|
|
The device code was updated.The investigation determined the customer's actions (replacing the reagent pack and forcing a new lot calibration) resolved the issue.
|
| |
|