• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY Back to Search Results
Catalog Number 08445699190
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2023
Type of Reportable Event Malfunction
Event or Problem Description
We received an allegation of questionable results for 1 patient sample tested with a1cx3 assay on a cobas pro c 503 analytical unit serial number (b)(4).The following results were reported: initial result: 10.8%, repeated result: 9.5%.The results were reported outside of the laboratory.The repeated result of 9.5% deemed to be correct.
 
Additional Manufacturer Narrative
The field service engineer found that a1c qc was out of range.The customer replaced the calibrator with a new lot.Qc was performed and it was within the normal range.
 
Additional Manufacturer Narrative
The device code was updated.The investigation determined the customer's actions (replacing the reagent pack and forcing a new lot calibration) resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TINA-QUANT HEMOGLOBIN A1CDX GEN. 3
Common Device Name
GLYCOSYLATED HEMOGLOBIN ASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16681835
Report Number1823260-2023-01096
Device Sequence Number8079851
Product Code LCP
UDI-Device Identifier07613336157824
UDI-Public07613336157824
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K072714
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number08445699190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2023
Supplement Date Received by Manufacturer04/12/2023
Initial Report FDA Received Date04/05/2023
Supplement Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1