|
SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 16MM; SCREW, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
04.211.016 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/10/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthese reports an event in japan as follows: it was reported that on (b)(6) 2023, the patient underwent the open reduction and internal fixation (orif) surgery for clavicular diaphyseal fracture with va-cp diaphysis cs2 implants.Procedure was completed successfully without any surgical delay.After 6 weeks, on (b)(6) 2023, of non-weight bearing, at the moment when the patient tried to lift the arm for the first time, 3 locking screws 2.7 mm in question on the proximal side were broken, and 1 cortical screw in question came off.Revision surgery was done on (b)(6) 2023, with our anteria clavicle implants.The revision procedure was completed successfully without any surgical delay.The surgeon commented that when using the 3.5 mm plate of the past standard, the screw did not break even if the reduction position was somewhat poor, but this time the screw broke because the new product standard was smaller, and the strength was weaker.No further information is available.This report is for one (1) 2.7mm ti va lacking scr self-tapng with t8 stardive recess 16mm.This is report 2 of 4 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Photo investigation: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device images.Visual analysis of the photo revealed that va lockscr ã¸2.7 he 2.4 self-tap l16 tan had the threaded head stripped.No other defect was found.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for va lockscr ã¸2.7 he 2.4 self-tap l16 tan.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|