• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 16MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 16MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 04.211.016
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthese reports an event in japan as follows: it was reported that on (b)(6) 2023, the patient underwent the open reduction and internal fixation (orif) surgery for clavicular diaphyseal fracture with va-cp diaphysis cs2 implants.Procedure was completed successfully without any surgical delay.After 6 weeks, on (b)(6) 2023, of non-weight bearing, at the moment when the patient tried to lift the arm for the first time, 3 locking screws 2.7 mm in question on the proximal side were broken, and 1 cortical screw in question came off.Revision surgery was done on (b)(6) 2023, with our anteria clavicle implants.The revision procedure was completed successfully without any surgical delay.The surgeon commented that when using the 3.5 mm plate of the past standard, the screw did not break even if the reduction position was somewhat poor, but this time the screw broke because the new product standard was smaller, and the strength was weaker.No further information is available.This report is for one (1) 2.7mm ti va lacking scr self-taping with t8 stardrive recess 16mm.This is report 1 of 4 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Photo investigation: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device image.Visual analysis of the photo revealed that va lockscr ã¸2.7 he 2.4 self-tap l16 tan had the threaded head stripped.No other defect was found.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for va lockscr ã¸2.7 he 2.4 self-tap l16 tan.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 16MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16682076
Report Number8030965-2023-04145
Device Sequence Number12754389
Product Code HWC
UDI-Device Identifier10886982119001
UDI-Public(01)10886982119001
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103243
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.211.016
Device Catalogue Number04.211.016TS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2023
Supplement Date Received by Manufacturer06/13/2023
Initial Report FDA Received Date04/05/2023
Supplement Report FDA Received Date06/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORTSCR ø2.7 SELF-TAP L18 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L16 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L18 TAN; VA-LCP CLAVIC PL 2.7 SHAFT CS2 LE TAN
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-