| Model Number |
Sterile Lighthandle Cover |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Steris has confirmed the reported issue.The polyblend plastic material that is used to form the light handle cover was sourced from a secondary supplier beginning in (b)(6)2022 and may not meet performance specifications.Steris initiated a recall (removal) of the affected product on (b)(6) 2023.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during patient procedures, light handle covers detached and fell into the sterile field.No report of injury.
|
| |
|
Search Alerts/Recalls
|