STERIS CORPORATION - MONTGOMERY LB53 LIGHT HANDLE COVER Back to Search Results
Model Number Sterile Lighthandle Cover
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Steris has confirmed the reported issue.The polyblend plastic material that is used to form the light handle cover was sourced from a secondary supplier beginning in october 2022 and may not meet performance specifications.Steris initiated a recall removal of the affected product on march 3, 2023.
 
Event or Problem Description
The user facility reported via medwatch report mw5115280 that during a patient procedure their light handle cover detached and fell into the sterile field.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LB53 LIGHT HANDLE COVER
Common Device Name
LIGHT HANDLE COVER
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key16683066
Report Number1043572-2023-00083
Device Sequence Number20656380
Product Code FTA
UDI-Device Identifier00724995150136
UDI-Public00724995150136
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSterile Lighthandle Cover
Device Catalogue NumberLB53
Device Lot Number9799082
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2023
Initial Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-