• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA FEM ADAPTER NEUTRAL BOLT; SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA FEM ADAPTER NEUTRAL BOLT; SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY Back to Search Results
Model Number 96-0783
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 03/26/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision completed due to knee pain and instability.Upon taking an x-ray, it was shown that the neutral adapter bolt had broken off and was floating in the patients knee.This meant the femoral component had come disengaged from the cemented sleeve and stem construct.Upon opening the knee, the broken bolt and femur were taken out easily.The 5 degree femoral adapter was still taper engaged in the 20mm cemented sleeve.There was loosening reported at the bone/cement interface.There was no surgical delay.Doi: unknown.Dor: (b)(6) 2023.Affected side: left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre), was not possible because the provided lot code was not valid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA FEM ADAPTER NEUTRAL BOLT
Common Device Name
SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16683075
Report Number1818910-2023-07433
Device Sequence Number7950180
Product Code JWH
UDI-Device Identifier10603295233930
UDI-Public10603295233930
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K060515
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-0783
Device Catalogue Number960783
Device Lot NumberD960783
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2023
Supplement Date Received by Manufacturer06/01/2023
Initial Report FDA Received Date04/05/2023
Supplement Report FDA Received Date06/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC*SIGMA TC3 FEM LEFT S2.5; SIG TIB CEM STM 13X60 2/2.5/3; SIGMA FEM ADAPTER 5 DEGREE; SIGMA FEM ADAPTER NEUTRAL BOLT; TC3 RP TIBIAL INSERT S2.5,12.5; UNIV FEM SLEEVE CEM 20MM; UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexFemale
-
-