It was reported that a minimum fill notification occurred after the user filled the cartridge with 200 units of insulin during the load sequence.Customer¿s blood glucose level was 522-523 mg/dl; cause was due to no insulin being delivered by pump.Reportedly, customer intentionally delivered insulin via fill tubing process while connected to the site to address bg level.The cartridge was reloaded to resolve the issue.Multiple attempts were made to follow up with the customer regarding the reported issue; however, no response from the customer was received.No additional patient or event information was available.
|