• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACCESS VASCULAR INC. HYDROPICC 5FR DUAL LUMEN; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACCESS VASCULAR INC. HYDROPICC 5FR DUAL LUMEN; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number PICC-251
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Type of Reportable Event Serious Injury
Event or Problem Description
Dvt reported for patient with picc-251 device.
 
Additional Manufacturer Narrative
On (b)(6) 2023, (b)(6) (avi clinical field director) and (b)(6)(avi national sales manager) were at (b)(6), a site participating in the picc-251 (dual lumen) limited market release (lmr).During this visit, a radiologist mentioned that one of the patients had a dvt and went on to state that "it was a difficult placement".For this patient, the vascular access team at riverside had attempted "multiple" times to insert a non-avi picc line catheter without success.The patient was then referred to the ir team to insert an avi picc-251 catheter; it took the ir team "multiple" tries as well but they were able to successfully insert the picc-251 line on (b)(6).Through (b)(6) discussion with the team at riverside, the general consensus was that the dvt was not caused by the product, but rather a vein that was "well used" or "over-manipulated." avi reviewed the picc-251 lmr insertion questionnaire that was completed for the device.In summary: - the clinician (jeffrey mendell) had prior experience with avi catheter kits (>5 insertions).- the patient had a co-morbidity: diabetes.- the catheter was inserted in the left basilic vein, however, the patient was identified as having strictures (v-18 wire could not be withdrawn; v-14 was used to pass a stricture).The clinician rated the "ease of advancement of the catheter" a 3/5 (average).- the clinician determined the device to be clinically acceptable.The device was discarded after removal, therefore not available for evaluation by avi.No patient outcome was noted.The lot number provided, #11423187, is assigned by argon medical (avi's approved supplier for kitting).This kit contained catheters from avi catheter lot #08182206.The lhr for these catheters was reviewed and there were no deviations or nonconformances found that could contribute to the complaint issue.The instructions for use (ls-068 rev c) for picc-251 includes a contraindication for "anatomical irregularities (structural or vascular) which may compromise catheter insertion or catheter care procedures." consensus from the team at (b)(6) was that the dvt was not caused by the product, but rather a vein with a known blockage that was identified prior to placement of the line.No further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROPICC 5FR DUAL LUMEN
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
ACCESS VASCULAR INC.
749 middlesex turnpike
billerica MA 01821
Manufacturer (Section G)
ACCESS VASCULAR INC.
749 middlesex turnpike
billerica MA 01821
Manufacturer Contact
brian hanley
749 middlesex turnpike
billerica, MA 01821
7815386594
MDR Report Key16685340
Report Number3015060232-2023-00005
Device Sequence Number20639528
Product Code LJS
UDI-Device Identifier00850030354044
UDI-Public(01)00850030354044
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K213550
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/23/2025
Device Model NumberPICC-251
Device Catalogue Number80002002
Device Lot Number11423187
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2023
Initial Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-