• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 SOLITAIRE FR3; CATHETER, THROMBUS RETRIEVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 SOLITAIRE FR3; CATHETER, THROMBUS RETRIEVER Back to Search Results
Model Number SFR3-6-40-10
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Coma (2417)
Date of Event 03/26/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the stent was stuck during the procedure during delivery.It was noted that the vessel was moderately tortuous.Resistance occurred in the middle section of the catheter.The reported device and any accessory devices were prepared as indicated in the ifu.The patient was undergoing surgery for treatment of an ischemic stroke of the internal carotid artery, c7.Iv tpa was not contraindicated.There were 2 passes with the device.It was noted the patient's vessel tortuosity was minimal.Stroke onset to reperfusion time was 2 hours.It was unknown if there were any patient symptoms or further complications were reported as a result of this event.Additional information received that the patient recovering from the procedure is in a light coma.The patient did not experience any symptoms related to the event.The catheter was a medtronic product rebar27 lot # b412012.The patient¿s baseline nihss score 27.The patient¿s post-thrombectomynihss score 20.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis: as found condition: the solitaire platinum revascularization device was returned for analysis within a shipping box, within a plastic bio-pouch, and within its introducer sheath.Damage location details: no bends or kinks were found with the pusher.No damages were found with the marker coil or stent attachment zone (proximal marker).The stent was found still attached to the pusher.The stent non-working and working length struts were found in good condition.Testing/analysis: the solitaire platinum revascularization device was pushed out from within its introducer sheath without issue.Conclusion: based on the device analysis and reported information, the customer¿s ¿resistance¿ report could not be confirmed.No defect was found with the returned solitaire platinum revascularization device that would have contributed to the event.Possible causes of ¿resistance¿ include use of an incompatible catheter, catheter damage, patient vessel tortuosity, or user does not maintain continuous flush.The rebar-027 catheter used in the event was not returned; therefore, an analysis could not be performed and ¿catheter damage¿ could not be ruled out as a potential cause.The rebar-027 catheter is compatible with the solitaire platinum revascularization device.Information regarding if a continuous flush was used was not reported; therefore, ¿user does not maintain continuous flush¿ could not be ruled out as a potential cause.It is possible the patient¿s ¿moderate¿ vessel tortuosity contributed to the event.However, the cause for the reported resistance could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLITAIRE FR3
Common Device Name
CATHETER, THROMBUS RETRIEVER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key16685497
Report Number2029214-2023-00593
Device Sequence Number20862100
Product Code NRY
UDI-Device Identifier00847536034792
UDI-Public00847536034792
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160641
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/07/2023
Device Model NumberSFR3-6-40-10
Device Catalogue NumberSFR3-6-40-10
Device Lot NumberB231220
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2023
Supplement Date Received by Manufacturer04/19/2023
05/16/2023
Initial Report FDA Received Date04/05/2023
Supplement Report FDA Received Date04/25/2023
05/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Other;
Patient SexUnknown
-
-