ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Model Number 1011338-30
Medical Device Problem Codes Mechanical Jam (2983); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified right internal carotid artery that was 99% stenosed.The lesion was pre-dilatated with a 6x20mm viatrac balloon at 20 atmospheres.The 6x30mm acculink had difficulty deploying as resistance was noted when pulling the handle slider; therefore, it was only partially deployed.The device was removed under fluoroscopy.A new unspecified stent was used to successfully complete the procedure.There was no adverse patient effects reported and no clinically significant delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported mechanical jam and the reported activation failure were able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during advancement inadvertent mishandling and/or interaction with the heavily calcified, 99% stenosed anatomy resulted in the noted device damages (wrinkled sheath, multiple shaft bends) thus compromising the device and ultimately resulting in the reported mechanical jam and the reported activation failure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Common Device Name
CAROTID STENT
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key16685987
Report Number2024168-2023-03543
Device Sequence Number8116340
Product Code NIM
UDI-Device Identifier08717648076220
UDI-Public08717648076220
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P040012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model Number1011338-30
Device Catalogue Number1011338-30
Device Lot Number1082561
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/14/2023
Supplement Date Received by Manufacturer05/12/2023
Initial Report FDA Received Date04/05/2023
Supplement Report FDA Received Date06/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
Patient Weight75 KG
-
-