The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.In addition, our technician confirmed that the angle wire fluid damage, the insertion flexible tube fluid damage, the segment fluid damage, the control body fluid damage, the down pulley wire fluid damage, the left pulley wire fluid damage, the ccd module with drive pcb fluid damage, the lcb (light carrying bundle) fluid damage, the light guide cable fluid damage, the ccd drive pcb fluid damage, the lg connector fluid damage, the control body corroded, the lg connector corroded, the insertion flexible tube crushed, the operation channel (primary) resistance, the suction channel kink, and the remote control buttons misconfigured; however, these defects are not the main cause and/or irrelevant to the alleged complaint.Based on the technical report, ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|