| Model Number |
MOSAIQ |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that during treatment there was a beam timer error.The bmdm was reset and the modality became unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported that during treatment there was a beam timer error.The bmdm (beam mu dose/display module) was reset and the modality became unknown.It was ascertained from the mosaiq log files that the treatment delivery was interrupted by beam timer errors.At this point 458.2 mus of the prescribed 687.7 mus was delivered.The customer informed that the user reset the bmdm, as they wanted to resume treatment, but the actual modality was not registering.The remaining mus were delivered at the end of the day (last fraction) and there was no mistreatment.The cause of why the actual value was displayed as 'none' for modality on the machine (and showed as 'unknown' in mosaiq as a results of this) whilst the prescribed values were set cannot be established due to the sdd logs no longer being available for investigation from the customer site.The customer has been advised to retain the sdd logs and provide them to elekta if this error reoccurs.
|
| |
|