| Model Number |
CAR-505 |
| Medical Device Problem Codes |
Fracture (1260); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cartridge was discarded and not available for evaluation.A device history record (dhr) review was conducted for the reported lot number and revealed the product was released for distribution having met quality and manufacturing specifications and requirements.The instructions for use states "make sure mated luer-connectors are secure but do not over-tighten, especially when connections are wet.".
|
| |
|
Event or Problem Description
|
|
A report was received on (b)(6) 2023 from the clinical nurse manager (nm) at a critical care facility regarding a patient with unspecified pathology, who stated the cartridge luer connector broke when attempting rinseback of a continuous renal replacement therapy (crrt) and blood could not be returned to the patient on (b)(6) 2023.Additional information was received on (b)(6) 2023 from the nm stating that there were no symptoms or medical intervention required and that patient continued therapy with the nxstage system using a new cartridge.
|
| |
|