| Model Number |
168-550 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left deflation and textured exchange due to concerns of the product.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed opening on anterior side assessed as fold flaw opening.¿ anxiety-product/procedure: unable to observe as it is a medical event not related to the device.Additional observations: ¿ observed creases on the device.¿ yellow particles observed in the surface of the shell.¿ white calcification observed in the surface of the shell.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional later reported " implant removal from textured to smooth due to the concerns of the product." and "any crease associated with hole.".
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|