| Model Number |
0700410105 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: short in cord when moved.The failure(s) identified in the investigation is consistent with the complaint record.Root cause: short in cable due to the sterilization cycles and/or bending, kinking, high force impact or other damage to the cable.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|