| Model Number |
PROV40 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Low Cardiac Output (2501)
|
| Date of Event |
03/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient underwent an on-pump, heparinized mini mitral valve repair with concomitant maze procedure and left atrial appendage exclusion.A prov40 clip was placed successfully.Post-operative heart catheterization indicated that the left anterior descending artery (lad) was occluded.Patient was returned to the operating room where the surgeon performed a left thoracotomy and removed the prov40 and placed a balloon pump.Heart catheterization the next day showed the lad had normal flow.The patient is doing well.There was no reported device malfunction, and the adverse event was the result of a procedural complication.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4): the device was not returned for investigation as there was no reported device malfunction.The device history record was reviewed for lot 112045.There is nothing in the device history record that would indicate that the devices were released with any non-conformances that would contribute to the complaint.
|
| |
|